Electroporation Tips

NEB 10-beta (NEB #C3020) Competent E.coli Strains are available as electrocompetent cells.

  • Electroporation cuvettes and microcentrifuge tubes should be pre-chilled on ice. 
  • Electrocompetent cells should be thawed on ice and suspended well by carefully flicking the tubes. 
  • Once DNA is added to the cells, electroporation can be carried out immediately. It is not necessary to incubate DNA with cells. The maximum recommended volume of a DNA solution to be added is 2.5 µl. Addition of a large volume of DNA decreases transformation efficiency. 
  • Contaminants such as salts and proteins can lower electroporation efficiency. Ideally, DNA for transformation should be purified and suspended in water or TE. Transformation efficiency is more than 10-fold lower for ligation mixtures than the control pUC19 plasmid due to the presence of ligase and salts. If used directly, ligation reactions should be heat-inactivated at 65°C for 20 min and then diluted 10-fold. For optimal results, spin columns are recommended for clean up of ligation reactions. 
  • Electroporation conditions vary with different cuvettes and electroporators. If you are using electroporators or cuvettes not specified in the protocol, you may need to optimize the electroporation conditions. Cuvettes with 1mm gap are recommended (e.g. BTX Model 610/613 and Bio-Rad #165-2089). Higher voltage is required for cuvettes with 2 mm gap. 
  • Arcing may occur due to high concentration of salts or air bubbles. 
  • It is essential to add recovery medium to the cells immediately after electroporation. One minute delay can cause a 3-fold reduction in efficiency. 
  • Cold and dry selection plates lead to lower transformation efficiency. Pre-warm plates at 37°C for 1 hour. Using 37°C pre-warmed recovery medium increases the efficiency by about 20%. 
  • Refreeze unused cells in a dry ice/ethanol bath for 5 min and then store at -80°C. Do not use liquid nitrogen. Additional freeze-thaw cycles result in lower transformation efficiency.